For sterilizer and washer-disinfector validation a standard pressure transducer is not adequate as the process temperature is frequently in excess of 80°C. The Ashcroft KXDHT transducer is designed for continuous operation at process temperature up to 150°C and is much better suited for the purpose.

It can be used directly connected to the machine with no cooling loop. Supplied with a fixed cable terminated with a plug suitable for the Series 825 TVI. Other plugs can also be supplied.

Ashcroft KXDHT Pressure Transducer Specifications

  • Part number KXDHT.
  • Pressure range -1 bar to 4 bar gauge.
  • Accuracy ±0.25%.
  • Span thermal drift ±0.03%/°C.
  • Repeatability 0.03% full scale
  • Median temperature range -30°C to 150°C.
  • Operating temperature range -30°C to 85°C.
  • Overpressure limit 10 bar.
  • Signal output 4-20mA.
  • Power supply 24VDC from Anville data logger.
  • Process connection G 1/4" male.
  • Process connection material Stainless steel 3 16Ti.
  • Connection 3m captive cable terminated with a plug to suit.


If you would like more information, please contact us to "Get On TRAQuE"

Ashcroft KXDHT Pressure Transducer
SKU Ashcroft_KXDHT_Pressure_Transducer
Brand Anville Instruments